Big Creek Reserve
Water Quality Chemical Data Procedures:
Calibrate monitor(s) and check batteries at least every 6 months.
Change pH wand fill solution every 2 months. The DO sensor should
always be stored with either distilled water or DO fill solution
in the cap over the tip. The pH meter should always be stored
with 7.0 buffer solution in the cap over the tip.
Locations of sites:
BE is at the Big Creek beach area, just below and downstream of
the Highway 1 arch bridge.
BC is at the large pool in the Big Creek fork, just before the
creek joins Devil's Creek.
DC is at the base of the trail to the confluence, on the south
bank of Devil's Creek, just above the footbridge and before Devil's
Creek joins with Big Creek.
Before taking measurements:
" Examine DO sensor tip for bubbles and corrosion.
For bubbles: with the meter and the sensor apart, unscrew the
sensor tip. Holding the membrane unit facing upward, fill to the
top with sensor refill solution (in "checkmate" case,
plastic dropper with snap-off tip). Tap the sides gently to help
dislodge bubbles. Holding the sensor vertical above the membrane
tip, screw the tip into the membrane. Invert the unit and examine
the membrane for bubbles again. If there are bubbles, repeat the
procedure.
For corrosion: unscrew the tip, scrub it with the buffer (white
porous strip in checkmate case) until unit is shiny. Refill tip
with solution as for bubbles, above.
" Polarize DO meter: With the meter off, plug DO sensor into
the meter for at least 3 hours, preferably overnight, before use.
" Check the pH kcl solution. If the sensor is filled up to
the black rubber plug near the top of the clear section, it's
ok. If it is low, refill it with kcl solution from the "checkmate"
case. Reclose the unit firmly with the black rubber plug.
To take measurements:
1. Place the thermometer in a shady location.
2. Write down location, time, date, tester, and weather data on
data sheet.
3. Take conductivity measurement: place tip in stream area where
there is a consistent, gentle flow and turn meter on. Write down
final readout before turning the meter off. REMEMBER TO TURN IT
OFF.
4. Attach the probe to the pH meter. Remove the wetting cap and
place it right side up in the "checkmate" case. Briefly
unplug the black rubber plug at the top of the sensor body. REPLACE
THE PLUG. Place the sensor tip in an area of the creek with a
gentle, consistent flow and turn the meter on. It will beep when
it is done. Write down the measurement before turning it off.
REMEMBER TO TURN IT OFF. Put the wetting cap back on the tip.
Unplug the sensor from the meter and replace it in the "checkmate"
case.
5. Read the air temperature and write it down.
6. Place the thermometer in the creek until the temperature has
stabilized and record the temperature on the data sheet.
7. Remove the tip from the DO sensor without twisting it. Check
for bubbles. Place the tip in the water and press the button labeled
"read," then press "mode." Check the display
to be sure that the measurement will be in mg/l. The reading will
stabilize and stop before the water temperature has an accurate
reading, so using the previously recorded water temperature as
a guide, continue to press "read" until the meter registers
a temperature of <.6*C greater than the true temperature. Write
down final DO measurement. Turn off the meter by pressing "mode."
Replace the wetting cap, adding solution if necessary. Leave the
sensor plugged into the meter until the last location has been
monitored.
8. Take a water sample for measuring turbidity later.
9. Make notes of anything unusual that might have affected the
measurements.
10. Repeat the above steps 1-9 at the next two locations.
11. Record the depth as you drive by the staff gauge. Binoculars
will be needed.
12. Take all 3 turbidity samples to the lab. Fill one plastic
tube from the LaMotte turbidity kit with distilled water or Whale
Point tap water. Fill the other with water from one of the creek
samples. Use the 50ml fill line for fairly clean creek water,
use the 25ml fill line for dirty water. If there is a difference
between the creek water and the other water, shake the turbidity
reagent and add .5ml to the distilled/lab water. Stir it with
the rod provided. If the altered water is cloudier than the creek
water, the turbidity measurement is <5JTU. If the water is
less cloudy, add another .5ml of reagent and calculate accordingly.
Record the measurement. Rinse both tubes. Repeat for next two
samples, being sure to shake the reagent every time it is used.
Rinse and dry the tubes when done. Dispose of leftover creek water
and rinse and dry the sample bottles. Put the empty bottles back
in the "checkmate" case.